Optional check source use, Theory of operation – Fluke Biomedical 451B User Manual

Page 23

Advertising
background image

Ion Chamber Survey Meter

Theory of Operation

13

Optional Check Source Use

Warning

To prevent personal injury, use care in the handling and
storage of the check source. The check source contains
radioactive materials used for the verification of the Product
operation.

For more testing use the check source Model 450UCS (Fluke part number:

3267644), a radioactive button source that provides a nominal reading of

0.4 mR/h (4 µSv/h).
To check the device, do the following:
1. Turn on and initialize the Survey Meter.
2. After the self-test completes, wait for 2 minutes.
3. Open the beta shield.
4. Place the check source flat against the end window.
5. Check for a nominal reading of 0.4 mR/h (4 μSv/h).

Theory of Operation

Warning

To prevent personal injury, use care in handling this assembly
during removal and installation. This Product contains lithium
cells with a potential voltage of 63 V on the battery assembly.

Caution

To prevent damage to the Product, do not open or disassemble

the Product. The high impedance circuits of the ion chamber

are easily contaminated with grease and dirt that produce

electrical leakage.

The Survey Meter is an air ionization chamber meter calibrated in exposure rate

units of roentgens/hour (or Sieverts/hour) for gamma and x-ray in the energy

range of 20 keV to 2 MeV through the sliding Phenolic shield. It can also be used

for x-ray in the energy range of 7 keV to 30 keV with the sliding shield open. The

Survey Meter responds to, but is not calibrated for, beta radiation, with the slide

open. Beta energies that can be measured are above 100 keV. The two thin

Mylar windows have a combined density thickness of 6.6 mg/cm2. The Survey

Meter also responds to alpha radiation above 7.5 MeV.
The liquid crystal display shows the radiation rate in digital and analog form with

the range multiplier values also showing on the scale. It is a lightweight electronic

device that requires the computational capabilities of a microprocessor to make it

operate. It functions in a multiplex mode called quadruplex. This mode uses four

backplanes to accommodate the 128 elements of the display.