6 beta sensitivity, slide open, 7 beta shield response – Fluke Biomedical 90-12 User Manual

Page 7

Advertising
background image

Introduction

Photon Energy Response

1

1-3

Figure 1-1. Photon Energy Response

The energy response curve represents data taken below a rate of 200 mR/hr. Changes in the actual
curve could occur at rates over 200 mR/hr.

1.6 Beta Sensitivity, Slide Open

Table 1-2 shows the beta sensitivity of the 90-12 probe to a variety of sources.

Table 1-2.

Beta Sensitivity

Nuclide

Particle

Energy Max (MeV)

90-12 Sensitivity

CPM/Ci

90

Sr Beta 2.245/.544 362,000

137

Cs Beta/Gamma 1.176/0.662

45,662

36

Cl Beta

0.714

92,827

99

Tc Beta

0.295

560

14

C Beta

0.158

0


These data were taken with the beta shield moved to expose the open window. For these measurements,
one of the two windows was pointed directly at the source.

1.7 Beta Shield Response

The beta shield has a total average attenuation of 1890 mg/cm

2

. This rejects all alpha particles, beta rays

with an energy of 2.5 MeV or less, and non-penetrating x-rays when the slide covers the GM tube. See
the energy response curves for open and closed slide positions.

















Figure 1-2. General View