2. specifications – Fluke Biomedical 07-424 User Manual

Page 6

Advertising
background image

Nuclear Associates 07-424
Operators Manual

1-2

1-2. Specifications

Table 1-1. Instrument Specifications

Feature Specifications

Density Range:

0-4.50

Response Time:

2 seconds or faster at optical density 3.00

Warm-Up Time:

10 seconds maximum

Accuracy:

+0.02 at density of 3.00 with 1 mm aperture

Density Subtraction:

Zero control permits density subtraction of up to 4.50 with 3 mm aperture.
Minus sign included on display to show negative relative densities.

Zero Control:

10-turn control permits exact zeroing with flashing minus sign indicating
correct zero to within 0.002.

Zero Drift:

Negligible

Spectral Response:

Centered at approximately 465 nanometers

Apertures Supplied:

3 mm, 2 mm and 1 mm dia. Aluminum

Throat:

210 mm (8.25 in.)

Illuminated Table Area:

120 x 140 mm (4.75 x 5.5 in.)

Detector: Silicon

diode

Detector Light Source:

White LED provides extremely long life with minimum spectral and intensity
degradation. Also reduce specimen heating to a minimum.


DESIGNED TO MEASURE DIFFUSE TRANSMISSION
DENSITY PER ANSI SPECIFICATION PH2-1 9-1959.

Readout:

Located on measurement arm, three L.E.D. digits with minus sign, 10 mm
(0.4 in.) high

Dimensions:

4.9 in. x 6.3 in. x 13.4 in

(H x W x D)

(125 mm x 160 mm x 340 mm)

Weight:

7.5 Ibs. (3.4 kg)

Shipping Weight:

10 Ibs. (4.3 kg)

Power:

Switch selectable, 115 VAC, 50/60 Hz, 15W 230 VAC

Unit available: order #07-424-220

Fuses:

For 115 VAC use 1.0 A, 250 V, AGC, Fast Acting

For 230 VAC use 0.5A, 250 V, AGC, Fast Acting

Operating Conditions:

10° C - 40° C (50° F to 104° F) Maximum 90% relative humidity, non-
condensing

Accessories Supplied:

Four #37 lamps, one extra o-ring light seal, one set of three apertures