Fluke Biomedical 07-706 User Manual

Page 9

Advertising
background image

Introduction

Introduction

1

1-5

Table 2 Effect of Tube Potential, Distance and Filtration on Air Exposure Rate at Panel of

Fluoroscopes*

Equivalent Total Aluminum Filtration

1 mm

2 mm 2.5 mm 3 mm 4 mm

Potential
kVp

Source to

Panel

Distance

cm inches

Roentgens per Milliampere Minute

70

30 12
38 15
46 18

5.3 2.7 2.2* 1.8 1.3
3.5 1.7 1.4

1.2 0.8

2.4 1.2 1.0 0.8 0.6

80

30 12

38 15
46 18

7.0 3.9 3.2* 2.6 2.0
4.6 2.5 2.1

1.7 1.3

3.2 1.8 1.4 1.2 0.9

90

30 12
38 15
46 18

9.0 5.2 4.3* 3.6 2.8
5.8 3.3 2.8

2.3 1.8

4.0 2.3 1.9 1.6 1.2

100

30 12
38 15
46 18

11.0

6.6 5.5* 4.7 3.7

7.0 4.2 3.5

3.0 2.3

4.9 2.9 2.5 2.1 1.6

110

30 12
38 15
46 18

13.1

8.0 6.8* 5.9 4.6

8.4 5.1 4.4

3.8 3.0

5.8 3.5 3.0 2.6 2.0

120

30 12
38 15
46 18

14.7

9.3 8.0* 7.0 5.5

9.5 6.0 5.1

4.5 3.6

6.5 4.1 3.6 3.1 2.5

130

38 15
46 18

-- 6.8

5.9

5.2 4.2

-- 4.7 4.1 3.6 2.9

140

38 15
46 18

-- 7.6

6.6

5.9 4.8

-- 5.3 4.6 4.1 3.3

150

38 15
46 18

-- 8.5

7.5

6.7 5.4

-- 5.8 5.2 4.6 3.7

* Typical exposure rates produced by equipment with medium length cables, derived from references (8) and (14)
by interpolation and extrapolation. Filtration includes that of the tabletop and the x-ray tube with its inherent and
added filter. As used above, panel means either panel or tabletop.


Section 3.1.2(a) When the fluoroscope is operated at 80 kVp, the exposure rate in air at the position
where the beam enters the patient shall not exceed 3.2 R/mA-min at 30 cm and should not exceed 2.1
R/mA-min at 38 cm. For other tube potentials, the exposure per unit charge shall not exceed the values
marked with an asterisk (*) or a dagger (

).