Use of MCA User Data Memory by Specific Applications

Applications can use the MCA user data memory as needed. However, two principles should never
be violated. Applications should always delete the user data of another applications if they are not
able to retain the data consistency. Applications must not write one of the already existing
application names to byte offset 16.

MCA user data memory contents of applications from GBS Elektronik GmbH

Byte offset 0

U235:

Enrichment constant 1

WinSPEC:

Index for the analysis program

WinSCAN:

Scan method and scan parameter

Byte offset 4

U235:

Enrichment constant 2

WinSPEC-A:

Repeat mode options

WinSCAN:

Grid lines offset and period

Byte offset 8

U235:

Enrichment constant 3

WinSPEC-A:

Repeat mode options

Byte offset 12

U235:

Enrichment constant 4

WinSPEC-A:

Repeat mode options

Byte offset 16

Application name: “U235”, “LENG”, “SPEC”, “MCS”, “RATE”, “UF6 ”, “WMCA” , “WMCS” , “WSPC”, “WSCN”,
“WUF6”, “W235”, “MCAE”, “OSCI”

Byte offset 20

HV Inhibit Mode: 0, 1, 2, -1 (used for automatic restart in applications for unattended measurements)

Byte offset 22

HV Preset value: 0 ... 3600 (used for automatic restart in applications for unattended measurements)

Byte offset 24

Power switches

Byte offset 28

Stabilization preset: auto/channel

Byte offset 30

Stabilization on/off

Byte offset 32

Repeat mode : 0, 1, 2, 3, 4 ,100 (100 means software repeat mode), (Bit 15: 0 = inactive, 1 = active)

Byte offset 34

Elapsed repeats (software mode)

Byte offset 36

HV inhibit mode: 0, 1, 2, -1 (used as preset values after the MCA has been powered on)

Byte offset 38

HV preset value: 0 ... 3600 (used as preset values after the MCA has been powered on)

Byte offset 40

Energy calibration: channel 3

1

Byte offset 44

Energy calibration: energy 3

Byte offset 48

Energy calibration: channel 1 or null

Byte offset 52

Energy calibration: channel 2 or null

Byte offset 56

Energy calibration: energy 1 or actual slope

Byte offset 60

Energy calibration: energy 2 or actual offset

Byte offset 64

1

st

ROI pair (ROI end)

2

Byte offset 66

1

st

ROI pair (ROI begin)

Byte offset 68

2

nd

ROI pair

...

...

Byte offset 124

16

th

ROI pair

Byte offset 128

U235/UF6:

Inspection description

WinMCS:

16 ROI pairs of the MCA spectrum ( similar the other ROIs )

1 Older applications have not supported 3 point energy calibration. Therefore, they have certainly not written a 3

rd

channel/energy

pair to the user data memory. In order to distinguish whether the user data memory contains a valid 3

rd

channel/energy pair, the

data is coded.

2 In applications with MCS spectrum the ROIs of the MCS spectrum are saved here. In WinMCS the ROIs of the MCA spectrum