GBS Elektronik MCA-166 Influence of High voltage on HPGe detector resolution User Manual

Page 2

Advertising
background image

0

1 00 0

2 00 0

3 0 00

40 00

B ia s V olta g e (V )

0

10 0 0

20 0 0

30 0 0

40 0 0

p

e

a

k

p

o

s

it

io

n

P h o to n E ne rgy

4 3 k eV

6 09 ke V

1 33 3 keV

Fig 2: Dependence of peak position (=signal amplitude) on bias

0

10 00

20 00

30 00

40 00

B ia s V olta ge (V )

0 .2

0.4

0.6

0 .8

1

E

ff

ic

ie

n

c

y

(

re

l.

t

o

m

a

x

.

b

ia

s

)

P h o to n E n e rg y

4 3 k e V

6 0 9 k e V

1 3 3 3 ke V

Fig 3: Dependence of detector efficiency on bias.

Conclusions
-The detector tested is only an example. Other detectors may behave different.
-It is worth the effort to test, if a 4 kV detector can also be operated at 3 kV

GBS -Elektronik GmbH, Bautzner Landstr. 22,
01454 Großerkmannsdorf / b. Dresden
Tel.: (0351) 217007-0 Fax: (0351) 217007-21
homepage: http://www.gbs-elektronik.de/
email: [email protected]