CMG-5T Compact

3. Adjust the vertical screw until the output voltage reads zero.

4. Repeat steps 2 and 3 for the N/S and E/W channels, (with the

meter connected across pins C/D and E/F respectively).

5. If necessary, continue to adjust each channel in turn until

consistent results are obtained on all three channels.

6. Repeat steps 2 to 5 using the high-gain outputs, if available and

refine the settings as far as possible. The high gain outputs for the
vertical, N/S and E/W channels appear across pins S/T, Z/a and

H/K respectively, as shown in section 5 on page 24.

7. Ensure that the “O”-ring seal on the adjuster screw cap is present

and intact and replace the cap firmly, to keep the instrument's

electronics protected from water and dust.

After the cover is installed, the accelerometer outputs may drift until the
system establishes temperature equilibrium with its environment and

the sensor settles down in its position. If required, the offset adjustment
can be repeated to achieve a better output offset. With experience, it

should be possible to reduce the output level to less than ± 1 mV.

2.4.2

Removing offsets using a terminal connection

Note: this method will only work for the instrument connected to the

sensor A port on a digitiser.

Open a terminal window for the required sensor and enter the
command:

CENTRE MONITOR

or

CENTER MONITOR