K-Patents PR-21-S User Manual

Page 46

Advertising
background image

40

PR-21-S instruction manual

Figure 5.7

A slope graph

5.4.2 Diagnostic values

The values at the left of the graph are used for diagnostic purposes:

−

CONC

is the final concentration value including Field calibration adjustment, see

Figure 6.12.

−

TEMP

, see Section 5.4.3.

−

CCD

gives the position of the shadow edge on CCD in %.

−

nD

is the refractive index value n

D

from the sensor.

−

CALC

is the calculated concentration value without Field calibration adjustment,

Section 6.6.3

−

QF

or Quality Factor is a value in the range -1000–100. It measures the image

sharpness, a typical good value is 50. A QF value below 0 usually indicates
prism coating.

−

LED

is a measure of the amount of light from the light source in %. Should be

below 100 %.

−

HD TMP

= sensor head temperature, see Section 5.4.3.

−

HD HUM

= sensor head humidity, see Section 5.4.4.

−

I_SNS

value shows the current to sensor, the nominal value is 40 mA.

−

DTR TMP

= Indicating transmitter temperature, see Section 5.4.3.

−

DTR V1

gives the voltage from the power module, the nominal value is 24V.

−

DTR V2

gives the DC supply voltage, the nominal value is 3.3 V.

Note:

The Slope display also has a soft key

SENSOR RESTART

which can be used to

restart the current sensor (see upper left corner of the display for sensor letter)
after a sensor software update.

5.4.3 Temperature measurement

The system contains three different temperature measurements displayed to the
left of the graphs in Figure 5.6:

TEMP

is the process temperature used for automatic temperature compensation in

the Indicating transmitter (Section 6.6, “Calibrating the concentration measure-
ment”).

HD TMP

measures the temperature on the Sensor processor card PR-10100 (Fig-

ure 10.2).

DTR TMP

measures the temperature on the Motherboard of the Indicating transmit-

ter (Figure 3.6, “The Motherboard of the Indicating transmitter”).

Advertising