CB-7021, CB-7022, CB-7024 User’s Manual
CB-7022 Voltage Output Calibration Sequence:
1.
Connect voltmeter to module’s voltage output.
2.
Warm-Up for 30 minutes.
3.
Set output type to 2. (0 to 10V)
-> Refer Sec. 2.24.
4.
Program output 10V.
-> Refer Sec. 2.15.
5.
Check the meter and trim the output for
10V match by using trim command.-> Refer Sec. 2.18.
6.
Perform 10V Calibration Command.
-> Refer Sec. 2.21.
7.
Repeat step 1 to 6 for channel 1.