Measurement Computing Personal Daq/3000 Series User Manual
Page 69
Advertising

Worst Case Temperature Measuerment Error vs. Personal Daq3000 Ambient Temperature
with Thermocouple at 0ºC (Excludes Thermocouple Error);
AutoZero Disabled; Selected Thermocouple Types: T, J, K, and E
0
1
2
3
4
5
6
7
8
-30
-20
-10
0
10
20
30
40
50
60
70
Ambient Temperature (ºC)
T
e
m
p
e
rature Mea
s
urem
e
n
t E
rro
r (
ºC
) f
o
r T
y
p
e
s T, J, K, E
J
K
T
E
Worst Case Temperature Measuerment Error vs. PDaq3000 Ambient Temperature
with Thermocouple at 0ºC (Excludes Thermocouple Error);
AutoZero Enabled; Selected Thermocouple Types: T, J, K, and E
0.0
0.5
1.0
1.5
2.0
2.5
3.0
-30
-20
-10
0
10
20
30
40
50
60
70
Ambient Temperature (ºC)
Te
mp
e
ra
tur
e
M
e
a
s
ur
e
m
e
n
t Er
ror (
ºC
) f
o
r Ty
pe
s
T, J
, K, E
J
K
T
E
Note
: Specifications are subject to change without notice.
937492
Specifications 7-3