Hardware setup, Wbk11a – specifications, Configuration – Measurement Computing WBK11A User Manual

Page 2: Installation

Advertising
background image

pg. 2, WBK11A

988396

WBK11A – SSH Card

Hardware Setup

Configuration

All WBK11A configurations are controlled by software. There are no hardware settings.

Reference Note:
For detailed WaveView information refer to the

WaveView Document Module

. A PDF version

of the document can be accessed from the data acquisition CD via the <View PDFs> button on
the CD’s opening screen.

Installation

There is no user installation permitted. See notes on page 1.

WBK11A – Specifications

Name/Function: WBK11A; 8-Channel Simultaneous Sample-and-Hold Card
Number of Channels: 8
Connectors: Internal to the WaveBook/512 series, or WaveBook/516 series device;

a 36 pin-socket connector mates with a 36-pin connector

Accuracy: ±0.025% FS for WaveBook/512A applications.

For WaveBook/516 Series applications, see the table on page 3 of this WBK11A document module.

Offset: ±1 LSB max
Aperture Uncertainty: 75 ps max
Voltage Droop: 0.1 mV/ms max
Maximum Signal Voltage: ±5.00 VDC (×1)

Input Voltage Ranges:
Before a scan sequence begins, the input voltage ranges can be programmed via software.
The ranges can be expanded as follows:

Unipolar:

Unipolar applies to WBK 10A only.

0 to +10 V
0 to +5 V
0 to +2 V
0 to +1 V
0 to +0.5 V
0 to +0.2 V
0 to +0.1 V

Bipolar: Bipolar applies to

WaveBook/516, /516A, /516E, /512A, and WBK10A

-10 to +10 V
-5 to +5 V
-2 to +2 V
-1 to +1 V
-0.5 to +0.5 V
-0.2 to +0.2 V
-0.1 to +0.1 V
-.05 to + .05 V

This range applies to WBK 10A only.

Programmable Gain Amplifier Gain Ranges: ×1, 2, 5, 10, 20, 50, 100

Weight: 0.14 kg (0.3 lb)