To calculate the effect of the droop rate on the accuracy of your measurement, the
range of the signal, the resolution of the A/D converter and the time between HOLD
and A/D acquisition are needed. Here is an example.

RANGE CALCULATION
Signal Range

±50 millivolts

Gain

100

Amplifier Output

±5 Volts

At full scale

RESOLUTION CALCULATION
A/D resolution

12 bits

1 part in 4096

Volts per bit

2.44 millivolts

10/4096

TIME TO SAMPLE CALCULATION
Number of channels

8

Sample Rate

4,000 samples/ sec.

Time between samples

250 microseconds

1/4000

Time from HOLD to CH8

2 milliseconds

0.000250 * 8

DROOP CALCULATION
Maximum droop to CH8

±0.0001V * 2 = ±2 microvolts.
Less than 8% of the value of 1 bit.

It is clear from this calculation that droop rate is not a problem even at low sampling
rates. The sample rate would have to drop to less than 400 samples/ sec. over eight
channels before the droop was detectable by a 12 bit A/D converter.