4 cable connection to converter – Tecfluid MX4 For FLOMID Series User Manual
Page 24
Advertising

24
Terminal
16
Top coil (red cable)
17
Bottom coil (black cable)
18
Rear electrode (white cable)
19 Ground
(shield)
20
Front electrode (black cable)
So that the flow direction shown in the instrument matches the actual direction, please take into
account the cables colour scheme according to previous figure.
Connect the two cable connectors in their corresponding male connectors on the sensor,
mating the bumps in the guide of the male connectors.
Slide the cover on the sensor neck until it meets its stop.
Tighten the two fixing screws to anchor the cover.
Tighten the cable gland to guarantee water tightness.
3.4
Cable connection to converter
REMOTE SENSOR