Part 4: Appendices

Reduced voltage operation

If the local voltage is below that specified on page 6, and there is a high
pumphead loading, provision has been made in the design of the drive for a
simple modification to be carried out. This will make the drive suitable for use
with voltages as low as 180 volts when the 220-240V setting is selected, and as
low as 90V when the 100-120V setting is selected.

Please note that the work should be carried out only by qualified personnel and
that there are components at mains voltage inside the drive. If the drive is found
to be damaged after this modification is incorporated and the damage is
considered to have been caused by connecting it to a "normal " voltage, the
warranty will be considered to have been voided.

The modification requires the connector J18 on the control PCB to be reversed.
To locate the terminal, isolate the mains supply then remove the pump cover
which is secured by 4 screws located on the underside of the pump. Lift the
cover to one side

Specification

Supply

100-120/220-240V~50/60Hz

(For 80/180V operation see Part 4: Reduced voltage operation)

Speed control range

220:1

Nominal maximum rotor speeds

55rpm or 220rpm

Power consumption

100VA

Shaft torque

2.2Nm

Operating temperature range

5C to 40C

Storage temperature range

-40C to 70C

Weight (including 501RL pumphead)

7.7kg (17lbs)

Noise

<70dBA at 1m

Standards

IEC 335-1, EN60529 (IP31)

Machinery directive

98/37/EC EN60204-1

Low voltage directive

73/23/EEC EN61010-1

EMC directive

89/336/EEC EN50081-1/EN50082-1