Widex B2-CIC User Manual

Page 15

Advertising
background image

28

29

Removing the used wax guard
Wipe.away.any.visible.wax.from.the.shell.of.your.
hearing.instrument ..The.used.wax.guard.is.in.the.
stainless.steel.bushing.around.the.sound.outlet .
With.the.grey.holder.perpendicular.to.the.sound.
outlet,.insert.the.removal.hook.into.the.used.wax.
guard.so.that.the.shaft.of.the.holder.is.touching.
the. rim. of. the. wax. guard,. then. slowly. pull. the.
wax.guard.

straight.out ..There.is.no.need.to.dig.

or.twist.the.wax.guard.from.the.bushing .
A .. Vent.opening
B .. Sound.outlet
C .. Used.wax.

guard

A.

B.

C.

Inserting a new wax guard
Now.turn.the.grey.holder.around.so.that.the.used.
wax.guard.is.away.from.the.instrument.and.the.
new. one. is. in. position. to. be. inserted. into. your.
hearing.instrument ..
Gently.press.the.new.wax.guard.straight.into.the.
opening.of.the.sound.outlet ..Do.not.force.the.wax.
guard.into.the.sound.outlet ..The.outer.ring.of.the.
wax. guard. should. sit. flat. on. the. stainless. steel.
bushing .

Advertising