Specification

Function

Surveying atmospheres for Mercury (Hg) concentrations below and
above the accepted exposure limit


Detector

Dual beam Ultra-Violet absorption module


Measuring Ranges

0.1-199.9 and 1.0-1999 µg/m

3

(user controlled)


Sensitivity

0.1 µg/m

3

and 1.0 µg/m

3


Accuracy

± 5 micrograms or ± 10% of reading


Repeatability

± 5% FSD


Response Time

Approximately 3 seconds


Temperature Range

+10°C to +50°C


Battery Type

15 Volt NiMH rechargeable


Battery Life

Greater than 6 hours after full charge


Dimensions

145 x 295 x 80 mm (120 mm with handle)


Weight

5 lb 3 oz (2.35 kg)


Linearity

Higher than 5% from 0 to 500 µg/m

3


Zero Drift

Less than 5 µg/hour


Alarms

Audible alarm factory preset to 20 µg/m

3


Datalogger

0-2 volt for use with a datalogger (not supplied)


Operation

After a short warm up the MVI gives real time indication of Mercury
vapour levels through the PTFE probe



















Advertising