World class 3000 – Emerson WORLD CLASS 3000 User Manual
Page 47

Instruction Manual
Appendix A Rev. 3.6
July 1998
Rosemount Analytical Inc. A Division of Emerson Process Management
Appendices A-13
World Class 3000
Table A-2. Fault Finding (Continued)
Symptom
Check
Fault
Remedy
2. System responds to
oxygen concentration
changes but does not
give correct indication
Recorder or remote
indicator
Calibration error
Recalibrate recorder or
indicator. Reference Re-
corder Instruction Manual.
System calibration
Calibration error
Recalibrate system.
Probe mounting and con-
dition of duct
Air ingress into duct
Stop air leaks or resite
probe.
Good response, with
incorrect indication
Cell mV input to
electronics
Failure of electronics
Check electronics
package.
Calibration gas input port
Blocked port
Clean port. If the flue gas
is condensing in the cali-
bration gas line, insulate
the back of the probe.
Make sure that the cali-
bration gas line is capped
between calibrations, or a
check valve is installed.
3. System does not give
accurate indication of
applied calibration gas
Ceramic diffusion element
Diffusion element cracked,
broken, or missing
Replace diffusion element.
4. System takes a long
time to return to flue
gas value after calibra-
tion gas is turned off
Diffusion element
Plugged diffusion element
Change diffusion element
or snubber diffusion
element.
A