Read internal coefficients (command 'u') – Pressure Systems 98RK-1 User Manual

Page 88

Advertising
background image

Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©

Page 86

www.PressureSystems.com

READ INTERNAL COEFFICIENTS (Command 'u')

Purpose:

Returns one (or more contiguous) requested internal coefficient(s) in a specified

internal coefficient array, and in the specified response data format.

Command

ufaacc[-cc]

‘u’ is the command letter.
f’ is the format field.
aa’ is the array index field.
cc[-cc]’ is coefficient index [or contiguous range].

Response

dddd[ dddd]...

dddd’ are the datum fields, each with a leading space.


Description: The 1-character format field (f) is a single digit that defines the format of each

returned datum in the response. All datum ( dddd) fields returned will be
preceded by a space character. Most coefficients have a floating point datum
type (f=0-1), while others have an integer datum type (f=5). Requesting an
improper format will result in an “N08” error response. Valid format types for
coefficients are shown in the following table:

f

Converts each internal value from..

max.char.

0

single binary float

to

7-10 digit signed decimal

[-xxx]x.xxxxxx”

13

1

single binary float

to

8-digit hex “ xxxxxxxx”

9

5

long binary integer

to

8-digit hex “ xxxxxxxx”

9


The

2-character

array index field (aa) is a hexadecimal value selecting a

particular internal coefficient array. The first array index (aa=01) refers to

channel 1's transducer, the 16th (aa=10) refers to channel 16's transducer.

Finally, the last array (aa=11) refers to a special global array.


The

2-character

coefficient index field (cc) is a hexadecimal value that selects a

particular coefficient within the specified array. Multiple contiguous coefficients

may be specified by using a coefficient index “range” specified by adding a

hyphen (negative sign) between two such indexes (cc-cc).