Energizer ATX612 User Manual

Page 14

Advertising
background image

22

23

I

TEM

P

ART

N

UMBER

D

ESCRIPTION

(Q

TY

)

1

1705-8868

Main PC Board (1)

2

1705-6755

Display PC Board (1)

3

1705-9247

Interface PC Board (1)

4

1705-9031

Chassis (1)

5

1706-3447

Pump Assembly (1)

6

1705-9494

Nicad Battery Pack

or

1705-9312

Alkaline Battery Pack

7

1706-0716

Case Gasket (1)

8

1706-2498

Filter Housing (1)

9

1705-8157

Dust Filter/Water Stop (1.2 Micron) (1)

10

1705-7118

Speaker (1)

11

1705-8660

Speaker Gasket (1)

12

1705-8140

Speaker Water Barrier (1)

13

1706-0674

Sensor Manifold (1)

14

1702-8374

External Alarm Jack (1)

15

1705-0295

RFI Screen (1)

16

1706-1284

Case Top (1)

17

1705-8306

Key Pad (1)

18

1706-0500

Alarm Lens (1)

19

1705-8652

Alarm Lens Gasket (1)

The following items numbers refer to the exploded view
drawing on pages 24 and 25.

9.

R

EPLACEMENT

P

ARTS

WARNING: When removing a sensor from

service, the appropriate sensor plug
must be installed to ensure proper
instrument operation. Toxic/O

2

Sensor Plug P/N 1704-6947
Combustible Sensor Plug
P/N 1706-2647

8.4.1 TOXIC/OXYGEN SENSOR REMOVAL/

REPLACEMENT

• To remove a toxic or oxygen sensor, grasp the sensor

and lift it straight up. Use care to avoid bending the
sensor pins.

WARNING: When removing a toxic sensor for

storage, connect a shorting wire to
the two pins as shown.

• New toxic sensors are shipped with a shorting wire

attached. Remove the shorting wire from the new
sensor.

• Immediately install the sensor in the instrument. Never

apply pressure to the area inside of the black O-Ring seal
at the top of the sensor.

8.4.2 COMBUSTIBLE SENSOR REMOVAL/

REPLACEMENT

• To remove the combustible sensor, grasp the sensor and

lift it straight up.

• Press the new sensor firmly into the sockets on the

PC board.

8.5 C

HANGING THE

I

NTERNAL

F

ILTER

If the internal dust filter/water stop becomes blocked, a
PUMP FAULT alarm will result. The filter must be
replaced immediately before continuing operation. To
remove and replace the internal dust filter/water stop from
the instrument:

• Unscrew and remove the knurled sample inlet fitting from

the instrument as shown in Figure 6.

• Grasp the dust filter/water stop from the instrument and

pull straight out to remove.

• Replace the dust filter/water stop making sure that the

larger diameter opening of the filter is facing inward.

• Replace the knurled sample inlet fitting.

NOTE: The ATX612 will remain in a PUMP FAULT

condition if operation is attempted with the dust
filter/water stop removed.

!

!