2 internal weight calibration, Excell precision co., ltd, Procedure – Excell Precision FM User Manual

Page 17: Zero calibration, Weight calibration

Advertising
background image

EXCELL PRECISION CO., LTD

.

FM User Manual

SME300000119

16

5-2 Internal Weight Calibration

Turn on and warm up the machine for 15 to 30 minutes before calibration.

Adjust calibration switch to

“ON”, and the screen will display

.

Press or to select

.

Procedure


Zero

Calibration



Weight
Calibration












Zero Calibration

a ) Ensure nothing on the platter; after being stable, press , and the screen will display

“

…….

”. The zero calibration will be complete 5 seconds later.

b ) To abort zero calibration, just press .


Weight Calibration

a ) Place an object, whose weight is known, on the platter, and input the weight value from

front panel. After being stable, press and the screen will display

“

…….

”. The

weight calibration will be complete 5 seconds later.

b ) To abort weight calibration, just press .

Zero calibration complete

Confirmed

Calibration

Switch

“OFF”

Place the

weight mass

and set the

weight

ENTER

ENTER

ENTER

ENTER

E S C

ENTER

ENTER

E S C

TARE

ZERO

E S C

E S C