Caution, Warning – Precision Medical Chrome Flowmeters User Manual

Page 3

Advertising
background image

CAUTION

• Flowmeters must be operated with the Flow Tube in a vertical,

upright position.

• Only personnel instructed and trained in its use should operate

this Flowmeter.

• Ensure all connections are tight and leak free.

• Only use oxygen-safe leak detector.

•

DO NOT autoclave.

•

DO NOT gas sterilize with EtO (Ethylene Oxide).

•

DO NOT clean with aromatic hydrocarbons.

•

DO NOT immerse Flowmeter in any kind of liquid. This will void

the warranty.

• The 1MFA3001, 4MFA1001 & 6MFA1001 Flowmeters may have a

factory installed restrictor. Prior to use, check Flowmeter labeling

for flow restrictions.

• The 1MFA3001, 4MFA1001 & 6MFA1001 Flowmeters contain a

glass Flow Tube which is fragile. Special care should be observed

to avoid breaking the Flow Tube.

(Continued on Inside)

WARNING

•

Use Flowmeters only for their “Intended Use” as described in this

manual.

• ALWAYS confirm prescribed flow before administering to patient

and monitor flow on a frequent basis.

• Flowmeters may contain magnetic, ferrous material that may

affect the results of an MRI.

To Reduce the Risk of Fire or Explosion:

•

ALWAYS follow ANSI and CGA standards for Medical Gas

Products and Flowmeters and Oxygen Handling.

•

DO NOT use or store oils, greases, organic lubricants or any

combustible materials on or near this Flowmeter.

•

DO NOT use near any type of flame or flammable/explosive

substances, vapors or atmosphere.

•

DO NOT smoke in an area where oxygen is being administered.