Component description, Item, Description a – Precision Medical PM5400 Series User Manual

Page 8

Advertising
background image

7

H

ELI

O

2

BLENDEr

HELIUM-OxygEn

COMPONENT dESCRIPTION

ITEM

DESCRIPTION

A

Oxygen Concentration dial

A dial used for selecting oxygen concentrations between 20%

-100% or 30%

-100%. The

F

I

O

2

scale is used for reference

only.

The actual

F

I

O

2

must be verified with an Alarmed

Oxygen Monitor / Analyzer.

This dial does not rotate 360°. The dial starts at 20% or

30% and ends at 100%.

B

Primary Outlet Port

A male DISS oxygen fitting with check valve that delivers flow

when engaged to any controlling device, such as a flowmeter.

C

Auxiliary Outlet Port

A male DISS oxygen fitting with check valve that delivers flow

when engaged to any controlling device, such as a flowmeter.

This outlet is equipped with a bleed valve that allows the user

to control if the bleed is ON or OFF. With the bleed in the “ON”

position, this outlet delivers accurate oxygen concentrations

in the following flows:

Model

Flow Range

High Flow

2 - 100 l/min

Low Flow

0 -

30 l/min

d

Auxiliary Bleed Collar

The collar is used to engage and disengage the bleed. The

bleed is necessary to maintain accurate

F

I

O

2

Concentration

below 15 l/min for the High Flow and ≤ 3 l/min for the Low

Flow. To activate the bleed, slide and rotate, (if applicable) the

knurled collar back until it contacts the cover. To deactivate the

bleed, pull and rotate, (if applicable) collar away from cover

until bleed flow valve is closed.

E

Oxygen Inlet Fitting

A female DISS or NIST oxygen fitting with one way valve that

is used to connect an oxygen supply hose.

F

Heliox Inlet Fitting

A male DISS or NIST heliox fitting with one way valve that

is used to connect a heliox supply hose.

G

Alarm

An audible alarm that sounds due to an excessive

pressure drop or deletion of either gas supply.

H

Rear Slide Mount with dove tail.