RIGOL
User’s Guide for DM3058/DM3058E
2-23
Table 2-8 Continuity Measurement Characteristics
Test Current
1mA
Range
Fixed at 2kΩ
Open-circuit Voltage
<8V
Input Protection
1000V (HI
terminal)
Beep Condition
0≤R
testing
≤Short-circuit impedance;
1Ω≤Short-circuit impedance≤2kΩ