Specifications

4

3

SPECIFICATIONS

Measuring ranges

30A / 300A / 3000A AC, switch selectable

Output sensitivity

100mV / 10mV / 1mV per A (AC coupled)

Accuracy (at 25ºC)

± 1% of range (45-65Hz)

Frequency range

10Hz to 20kHz (-1dB)

Phase error

<± 1° (45-65Hz), ± 10° (at 20kHz)

Linearity

± 0.2% of reading from 10% ... 100% of range

Position sensitivity

± 2% of range with cable > 25mm (1") from the coupling

External field

± 1% of range with cable >200mm (8") from the head

Minimum load

100k

for specified accuracy

Noise

8mV rms (0.3% of range) in 30A range
2mV rms (0.1% of range) in 300A and 3000A ranges

Gain variation

± 0.08% /

°

C

Operating temp.

0°C to 70°C (32

°

F to 158°F) electronics

Storage temp.

-20°C to +70°C (-4

°

F to 158°F)

Power supply

Two AA MN 1500 LR6 alkaline batteries or equivalent

Battery life

200 hours typical

Low battery

Indicated by a red LED

Enclosure

IP5X flame retardant UL94-VO rated

Material

Valox 357 X