Congratulations! You are now the owner of an AMPROBE Instrument. It has been

quality crafted according to quality standards and contains quality components and
workmanship. This instrument has been inspected for proper operation of all of its

functions. It has been tested by qualified factory technicians according to the
long-established standards of AMPROBE.

1) SAFETY
This manual contains information and warnings that must be followed for operating the

instrument safely and maintaining the instrument in a safe operating condition. If the
instrument is used in a manner not specified by the manufacturer, the protection
provided by the instrument may be impaired.

The meter meets the requirements for double insulation to IEC61010-2-032 (1994),
EN61010-2-032 (1995), UL3111-2-032 (1999).

ACD-10 PRO/TRMS PRO

Category II 600 Volts ac and dc.
Category III 300 Volts ac and dc.


PER IEC61010 OVERVOLTAGE INSTALLATION CATEGORY

OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be
supplied from the fixed installation.
Note – Examples include household, office, and laboratory appliances.

OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.
Note – Examples include switches in the fixed installation and some equipment for
industrial use with permanent connection to the fixed installation.


OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.
Note – Examples include electricity meters and primary over-current protection
equipment.