Range

frequency

Accuracy

1) 2)

40.00A, 400.0A, 1000A

50 Hz / 60 Hz

± (0.5% rdg + 5d)

40.00A, 400.0A

45 to 50 Hz, 60 to 500 Hz

± (2.0% rdg + 5d)

1000A

± (2.5% rdg + 5d)

40.00A, 400.0A

500 Hz to 3.1 kHz

± (2.5% rdg + 5d)

1000A

± (3.0% rdg + 5d)

ACA AutoVA

TM

Threshold: 1A AC (40hz ~ 500hz only) nominal

Crest factor:
40.00A & 400.0A:

< 2.5 : 1 at full scale; < 5.0 : 1 at half scale

1000A:

< 1.4 : 1 at full scale; < 2.8 : 1 at half scale

1)

induced error from adjacent current-carrying conductor: < 0.06A/A

2)

Specified accuracy is from 1% rdg to 100% rdg of range and for measurements made at

the jaw center. when the conductor is not positioned at the jaw center, position errors
introduced are:

Add + 1% rdg to specified accuracy for measurements made wiThin jaw marking lines
(away from jaw opening)
Add + 4% rdg to specified accuracy for measurements made BeYonD jaw marking lines
(toward jaws opening)

temperature

Range

Accuracy

-50°C to -20°C

± (2.0% rdg + 6°C)

-20°C to 300°C

± (2.0% rdg + 3°C)

-58°f to -4°f

± (2.0% rdg + 12°f)

-4°f to 572°f

± (2.0% rdg + 6°f)

Type-K thermocouple range & accuracy not included

frequency
Range

5.00 hz to 500.0 hz

Accuracy:

± (0.5% rdg +4d

Range

Sensitivity (Sine RMS)

40A

> 4A

400A

> 40A

1000A

> 400A

600V

> 30V

tHD% rdg-f

1)

(ACD-41PQ only)