Amprobe AM-250 Digital-Multimeter User Manual

Page 17

Advertising
background image

11

Max Hold (Voltage & Current)

Specified accuracy ± 50 digits for changes > 25ms in duration

DC Current

Range

Accuracy

Burden Voltage

400.0µA

2.0% + 5d

0.15mV/µA

4000µA

1.2% + 3d

0.15mV/µA

40.00mA

2.0% + 5d

3.3mV/mA

400.0mA

1.2% + 3d

3.3mV/mA

4.000A

2.0% + 5d

0.03V/A

10.00A*

1.2% + 3d

0.03V/A

*10A continuous, 20A for 30 second max with 5 minutes cool down interval

AC Current

Range

Accuracy*

Burden Voltage

50Hz -- 500Hz
400.0µA

2.0% + 6d

0.15mV/µA

4000µA

1.5% + 4d

0.15mV/µA

40.00mA

2.0% + 6d

3.3mV/mA

400.0mA

1.7% + 4d

3.3mV/mA

4.000A

2.0% + 6d

0.03V/A

10.00A*

1.8% + 4d

0.03V/A

*10A continuous, 20A for 30 second max with 5 minutes cool down interval

Ohms

Range

Accuracy

400.0e

0.8% + 6d

4.000ke, 40.00ke, 400.0ke

0.6% + 4d

4.000Me

1.0% + 4d

40.00Me

2.0% + 4d

Open Circuit Voltage : 0.4VDC typical
Audible Continuity Tester
Audible threshold : between 10e and 120e

Advertising