Amprobe SM-10 Sound-Level-Meters User Manual

Page 8

Advertising
background image

7

when LCd blinked, release the A/C and the hoLd buttons. The sound level

4.

meter will display 94.0 dBA . The calibration is done.
The 94 dB calibration process can be repeated until the meter reads 94.0

5.

dB. Recalibration cycle: 1 year.

SPECIfICAtIoNS
general

display:

4 digits LCd

Microphone:

1/2 inch electret condenser microphone.

Low Battery indication:

Replace battery when LCd displays “”.

power Supply:

9V nedA 1604, ieC 6f22, JiS 006p battery.

power Life :

Approx. 50 hrs (alkaline Battery).

Auto power off:

Approx. 5 min

environment:

indoor operation, < 2000 m

Temperature / humidity
operation:

5 °C to 40 °C (41°f to 104°f); < 80% Rh

Storage:

-10 °C to 60 °C (14°f to 140°f); < 70 %Rh

dimension:

200 x 55 x 38 mm (7.8 x 2.2 x 1.5 in)

weight:

233g. (0.5 lb.) including battery

-eMC: en61326-1 This product complies with requirements of the

following european Community directives: 89/336/eeC (electromagnetic
Compatibility) and 73/23/eeC (Low Voltage) as amended by 93/68/eeC (Ce
Marking). however, electrical noise or intense electromagnetic fields in the
vicinity of the equipment may disturb the measurement circuit. Measuring
instruments will also respond to unwanted signals that may be present within
the measurement circuit. Users should exercise care and take appropriate
precautions to avoid misleading results when making measurements in the
presence of electronic interference.
Electrical (Audio)

Standard applied:

ieC 651 Type2, AnSi 1.4 Type 2.

dynamic range:

50 dB.

Resolution:

0.1 dB, display Update: 0.5 sec.

Time weighting:

fAST(125mS), SLow(1 sec).

frequency range:

31.5 hz to 8 khz.

Measuring level range:

(Auto Range)

A weighting:

30 to 130 dB.

C weighting:

35 to 130 dB.

Accuracy:

± 1.5 dB.

Calibration cycle:

1 year.

Advertising