Amprobe TH-1 Relative-Humidity-Temperature-Probe-Style-Meter User Manual

Page 7

Advertising
background image

6

bATTERy REPLACEMENT

Turn off the meter. Use a coin to open the battery cover.

1.

Put the battery (CR2032) with the ”+” faces up.

2.

Replace the cover.

3.

TRoUbLESHooTINg
POWER ON BUT NO DISPLAY

Make sure you press the ”ON/ OFF” key for more than 100mS.

Check the battery conditions and replace if necessary


DISPLAY DISAPPEARS

Check for low battery indicator before the display is off. If the low

battery indicator appears, replace the battery.
Turn on the meter by pressing “ON/OFF+HOLD” keys to disable

Auto Power Off.

E02-Out of the lowest limit.
E03-Out of the highest limit.
E04-Check error message of the original data.
E11-Calibration required. The humidity sensor accuracy can not be
calibrated by the user due to special procedures required. Please
contact a qualified calibration house for service.
E33-Return the meter for repair

SPECIfICATIoNS

Size

17.5 Ч 4.2 Ч 1.6cm (6.89 x 1.65 x 0.63 in)

Battery

150 Hours

Power consumption

< 2.0mA

Sensor

Electronic Capacitance Polymer Film Sensor

Range:

Relative Humidity 0 to 100%
Temperature

-20 to 50°C (-4 to 122°F)

DP/WB

-20 to 59.9°C / -5 to 59.9°C

(-4 to 139.8°F / 23 to 139.8°F)

Accuracy:

Relative Humidity ±3% from 10 to 95%RH and ±5% for other

RH range @ 25°C / 77°F

Temperature

±0.6°C / ±1°F

Resolution

0.1%, 0.1°C / 0.1°F

The meter conforms to the following standards:
EN 50081-1/1992 : EN 55022
EN 50082-1/1997 : EN 55024
(EN 61000-4-2/-3/-8, ENV 50204)