Amprobe TM45A Digital-Thermometer User Manual

Page 8

Advertising
background image

4

Electrical Specifications

Measurement range:

°C mode: -50.0°C to 199.9°C; 0.1°

resolution -50°C to 1300°C; 1° resolution

°F mode: -58.0°F to 199.9°F; 0.1°

resolution -50°F to 1999°F; 1° resolution
Resolution, selectable: 0.1° or 1°
Accuracy (specified for operating temperatures over the

range 18°C to 28°C / 64°F to 83°F, for 1 year, not including
thermocouple error)

°C mode:

-50°C to 199.9°C: ±(0.3%rdg +1°C)
-50°C to 1000°C: ±(0.3%rdg +1°C)
1000°C to 1300°C: ±(0.5%rdg +1°C)

°F mode:

-58°F to 199.9°F: ±(0.3%rdg +2°F)
58°F to 1832°F: ±(0.3%rdg +2°F)
1833°F to 1999°F: ±(0.5%rdg +2°F)

Temperature coefficient: 0.1 times the applicable accuracy

specification per °C from 0°C to 18°C and 28°C to 50°C

Overload protection: 24V AC or DC

Maintenance

In case you do not obtain the expected measurement result:
1. Check the batteries.
2. Review the operating instructions for possible mistakes in

operating procedure.

3. Inspect the thermocouple probe for a broken or intermittent

connection.

Except for the replacement of the batteries, repair of the
thermometer should be performed only by the Factory or by a
Factory Authorized Service Center.
The front panel and case can be cleaned with a mild solution
of detergent and water. Apply sparingly with a soft cloth and
allow to dry completely before using.