Calibration failure

Check whether the low battery indicator displayed before calibration. if yes, replace the

batteries and try again.
Make sure the meter probe is well sealed when plugging into the salt bottle.

Make sure the free air temperature is within 23

+2 C and the airflow is stable.

E 11

RH calibration error. Re-calibrate again.

E 33

Circuit error in RH measurement. Return the meter for repair.

E 34

Circuit error in Temperature measurement channel. Return the meter for repair.


Note: When the meter is on, plug with a probe will also result in ”e 33¡ or e 34¡in Max. and
Min. mode.

SpECIFICATIoNS
Range:

RH measurement: 0~100%

Temperature: -10~60 °C (14 to140°f)

Dp :-73.4 to 59.9 °C (-100.0 to139.8°f)

WbT: -13.4 to 59.9°C (7.88 to 139.8°f)


Accuracy:

RH : +/-3% from 10-90% (at calibration temperature); +/-5% for others .

Temperature: +/-0.6

°C(+/-1°f)

Storage Temperature: -20°C to 60°C (-4°f to 140°f)

operating Temperature: 0°C to 50°C (32°f to 122°f)