Amprobe THWD-5 Relative-Humidity-Temperature-Meter User Manual

Page 8

Advertising
background image

7

Calibration failure

Check whether the low battery indicator displayed before calibration. if yes, replace the

•

batteries and try again.
Make sure the meter probe is well sealed when plugging into the salt bottle.

•

Make sure the free air temperature is within 23

•

+2 C and the airflow is stable.

E 11

RH calibration error. Re-calibrate again.

•

E 33

Circuit error in RH measurement. Return the meter for repair.

•

E 34

Circuit error in Temperature measurement channel. Return the meter for repair.

•
Note: When the meter is on, plug with a probe will also result in ”e 33¡ or e 34¡in Max. and
Min. mode.

SpECIFICATIoNS
Range:

RH measurement: 0~100%

•

Temperature: -10~60 °C (14 to140°f)

•

Dp :-73.4 to 59.9 °C (-100.0 to139.8°f)

•

WbT: -13.4 to 59.9°C (7.88 to 139.8°f)

•
Accuracy:

RH : +/-3% from 10-90% (at calibration temperature); +/-5% for others .

•

Temperature: +/-0.6

•

°C(+/-1°f)

Storage Temperature: -20°C to 60°C (-4°f to 140°f)

operating Temperature: 0°C to 50°C (32°f to 122°f)

Standard accessories

1 ea THWD-5 meter

•

1 ea probe on extension cable

•

1 ea Users Manual

•

1 ea Hard carrying case

•

4 ea batteries (AAA)

•

Advertising