To SET CH/ID to 00,00, press the “Hi/Lo LIMITS” key and
“ ” power key for more than 6 seconds with the meter
powered down. The meter will set channel and ID to
00,00 status. The second display will show 00, which
means that the channel and ID has been set to 00.

To check the channel and ID of the meter
When the meter is off, press “°C/°F” key and “ ” for 5
seconds, LCD’s main display will show channel number,
the second display will show ID number.

SPECIFICATION
Electrical

Temperature
Scale

Celsius or Fahrenheit user-selectable

Measurement
Range

J-Type -200°C to 1050°C,
(-328°F to 1922°F)
K-Type -200°C to 1370°C,
(-328°F to 2498°F)

Resolution

0.1°C or 0.2°F

Accuracy

Accuracy is specified for operating
temperatures over the range of 18°C
to 28°C (64°F to 82°F), for 1 year, not
including thermocouple error.
±(0.05%rdg+0.3°C) on -50°C to 1370°C
±(0.05%rdg+0.7°C) on -50°C to -200°C
±(0.05%rdg+0.6°F) on -58°F to 2498°F
±(0.05%rdg+1.4°F) on -58°F to -328°F

Temperature
Coefficient

0.1 times the applicable accuracy
specification per °C from 0°C to 18°C
and 28°C to 50°C
(32°F to 64°F and 82°F to 122°F).