Environmental limits for R-Series sensors

Figure 1-1:

Maximum process temperature in °F (°C)

176 (80)

140 (60)

104 (40)

68 (20)

32 (0)

–4 (–20)

–40 (–40)

–148 (–100)

158

(70)

Ambient temperature of core processor or

transmitter in °F (°C)

–112 (–80)

–76 (–60)

–40 (–40)

–4 (–20)

32 (0)

68 (20)

104 (40)

140 (60)

176 (80)

212 (100)

248 (120)

302 (150)

104
(40)

Notes

•

When ambient temperature is below –40 °F (–40 °C), a core processor or transmitter must be
heated to bring its local ambient temperature to between –40 °F (–40 °C) and +140 °F (+60 °C).
Long-term storage of electronics at ambient temperatures below –40 °F (–40 °C) is not
recommended.

•

Temperature limits may be further restricted by hazardous area approvals. Refer to the hazardous
area approvals documentation shipped with the sensor or available from the Micro Motion web site
(

www.micromotion.com

).

•

The extended-electronics option allows the sensor case to be insulated without covering the
transmitter or core processor, but does not affect temperature ratings.