World class 3000, Instruction manual – Emerson ROSEMOUNT 3000 User Manual

Page 53

Advertising
background image

Instruction Manual

IB-106-300NH Rev. 4.3

May 2005

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-17

World Class 3000

+

+

+

+

+

+

+

+

+

-

-

-

-

-

-

-

-

-

BK WH

J9

J8

J3

J2

J1

TRIAC RELAY

STACK

TC

ANALOG

HEATER

PROBE

MV

PROBE

MV

PROBE

TC

PROBE

TC

AD590

(INTERNAL

WIRING)

WH

BK

OR CELL+

YE HTR TC +

BL CELL -

RD HTR TC

-

BK

WH

GN

STACK

TC

PROBE

HEATER

R H

N

L

LINE
VOLTAGE

GN/YE

TOP

BOTTOM

4 TWISTED PAIR SHIELDED
#22 AWG BY CUSTOMER

2 TWISTED PAIR SHIELDED
#22 AWG BY CUSTOMER
(OPTIONAL)

HPS 3000 INTERFACE MODULE

LINE

1

2

2

7

11

1

4

WORLD CLASS PROBE

ALWAYS DISCONNECT LINE VOLTAGE

FROM HEATER POWER SUPPLY

BEFORE CHANGING JUMPERS.

JUMPER

CONFIGURATION

LINE VOLTAGE

SELECTION

JUMPER

(INSTALL)

JUMPER

(INSTALL)

HEATER

POWER

ON

120 V.A.C./100 V.A.C.

220/240 V.A.C.

JM4, JM1

JM5

REMOTE

ON

REMOVE JM2

INSTALL JM2

ELECTRONICS

SELECTION

JUMPER

PROBE HEATER

VOLTAGE SECTION

REMOVE JM3, JM6

DIGITAL

(NEXT GENERATION)

WORLD CLASS PROBE

JM7

NOTES:

1 2 3 4 5 6

7 8

BL

PROBE

MV

-

PROBE

MV

+

PROBE

TC

+

PROBE

TC

-

OR

YE

RD

GN

E

R

H

WH

BK

GN

CELL

-VE

OR

CELL

+VE

YE

CHROMEL

RD

ALUMEL

GN

BK

BK

}

HEA

TER

PROBE JUNCTION

BOX WIRING

A

B

37840007

RELAY WIRE IS OPTIONAL, RELAY CAN BE BYPASSED
WITH JUMPER JM-2 IF NOT WIRED TO THE IFT.

1

2

STACK TC WIRING AS REQUIRED.

3

ALL WIRES #16-#22 AWG TWISTED PAIR WITH SHIELD
BY CUSTOMER EXCEPT AS NOTED.

5

REMOVE JM1 ON INTERCONNECT BOARD.

6

INSTALL JM6 ON MICROPROCESSOR
BOARD PER FIGURE 2-8.

7

IF RELAY WIRE OF NOTE 1 INSTALLED THEN
REMOVE JM2 ON HPS 3000.

9

IF MPS 3000 NOT USED.

10

1 RELAY PER PROBE AVAILABLE FOR CALIBRATION
STATUS INDICATION (48 V max, 100 mA max).

11

100 V.A.C. REQUIRES TRANSFORMER PART
NUMBER 1M02961G02.

4

STANDARD PROBE CABLE BETWEEN PROBE AND
HPS BY ROSEMOUNT ANALYTICAL.

8

IF STACK TEMPERATURE NOT USED.

Figure 2-13. Wiring Layout for Complete IFT 3000 System with HPS (Sheet 1of 2)