Xdt user’s manual 30, Display explanation required action, Display is dark – Epson XDT User Manual

Page 37

Advertising
background image

XDT User’s Manual

30

Legend: 

denotes a beeping accompaniment to the message.

/

denotes alternately flashing messages.

DISPLAY

EXPLANATION

REQUIRED ACTION

LO ERR



PROM check sum failed.

power

on tes

ts

HI ERR



RAM write/read test failed.

cycle power

% ERR



Unidentified power-up failure.

if problem persists,

C.S.F/



EEPROM Check Sum Failed.

return to your representative

ERR/AD



A/D converter failure.

s

y

st

em

tes

ts

once p

er

2

mi

n.

for service.

ERR/REF



Reference voltage for A/D out of spec.

LO/BAT



Low supply voltage.

LO RNG/



TEMP

Instrument low temperature range

has been exceeded.

make sure that the unit is at a

HI RNG/



TEMP

Instrument high temperature range

has been exceeded.

temperature of -10°C to +50°C.

OPN/



Sensor circuit is open.

dewpoint dis

p

lay

te

st

s

SAT/



Sensor is saturated.

see table in sensor troubleshooting section 2.5

SHR/



Sensor circuit is shorted.

...

Trying to calculate dewpoint for undefined sensor.

select sensor and recalibrate.

DB



Trying to calibrate an undefined sensor.

cal

ib

ra

ti

o

n

te

st

s

see calibration instructions.

SEN/2LO



Sensor reading is ‘too’ low to be from a saturated sensor,

thus can not be calibrated

2LO

While Compensating for cable length, the measurement did not

correspond to the expected open or short values.

cabl

e

te

st

s

Check the means of opening & shorting the sensor

cable. see section 3.4.5 - 11

2HI

ERR/EEP



EEPROM write cycle not completed.

m

is

cel

laneous

if this persists, return for service.

LOC



Attempting to modify a locked unit.

unlock unit, see set-up state 3.4.5.-10.

SEC



Attempting to unlock a secured unit.

HI (flashing while
viewing DEWPOINT)

The HI Alarm (alarm #1) relay is deenergized.

See section 3.4.4.2 Setting the Alarms.

LO (flashing while
viewing DEWPOINT)

The LO Alarm (alarm #2) relay is deenergized.

A.O./OPN



The Analog Output ‘sees’ an open circuit

check the cables connecting the analog output, make

sure the terminating impedance is less that 500

Ω.

PSI flashing when dis-
playing dewpoint

Pressure correction is enabled, the dewpoint is computed using

programmed pressures

refer to section 3.4.4.1

-70°C (-94°F) with
flashing DEWPOINT

An XTR-65 sensor is measuring less than -70°C or -94°F D.P.

check instrument

upgrade to XTR-100 sensor

RNG

number can not be displayed in 3½ digits

XEN



Turn on message, Xentaur (Greek

Ξ

=X)

display is dark

No power to instrument

check power to instrument, check fuse