SC Series Instruction Manual page 117

Hig

h-L

ev

el

In

pu

t C

han

nel

12

12.6.6

CALIBRATE

Menu

Item

This menu item performs a calibration according to what was entered in the

CALI-

BRATION TYPE

and

CALIBRATION DATA

menu items.

If the

CALIBRATION TYPE

is Shunt Calibration...

•

The display will read

DOING SHUNT CAL

, and prompt you to

APPLY 00000 UNIT

(where

00000 and UNIT are previously entered as

ZERO SCALE VALUE

and

DISPLAY. UNITS

.

When you have applied this load to the transducer, press [ENTER].

•

The display will then read

WORKING

and calibration will take place. Upon com-

pletion, the display will indicate

DONE

and the instrument will return to the RUN

mode.

If the

CALIBRATION TYPE

is 2-Point Known Load Calibration...

•

The display will read

DOING 2POINT CAL

, and prompt you to

APPLY 00000 UNIT

(where 00000 and UNIT are previously entered as

KNOWN POINT 1/2

and

DISPLAY.

UNITS

). When you have applied this load to the transducer, press [ENTER].

•

The display will read

WORKING

, then

APPLY 01000 UNIT

(where 01000 and UNIT are

previously entered as

KNOWN POINT 2/2

and

DISPLAY. UNITS

). When you have

applied this load to the transducer, press [ENTER].

•

Upon completion, the display will indicate

DONE

and the instrument will return to

the RUN mode.

If the

CALIBRATION TYPE

is 3-Point Known Load Calibration...

•

The display will read

DOING 3POINT CAL

, and prompt you to

APPLY 00000 UNIT

(where 00000 and UNIT are previously entered as

KNOWN POINT 1/3

and

DISPLAY.

UNITS

). When you have applied this load to the transducer, press [ENTER].

•

The display will read

WORKING

, then

APPLY 00500 UNIT

(where 00500 and UNIT are

previously entered as

KNOWN POINT 2/3

and

DISPLAY. UNITS

). When you have

applied this load to the transducer, press [ENTER].

•

The display will read

WORKING

, then

APPLY 01000 UNIT

(where 01000 and UNIT are

previously entered as

KNOWN POINT 3/3

and

DISPLAY. UNITS

). When you have

applied this load to the transducer, press [ENTER].

•

Upon completion, the display will indicate

DONE

and the instrument will return to

the RUN mode.

NOTICE

Before performing a calibration,
•

the transducer must be connected to the instrument,

•

the channel’s Excitation and Signal Jumpers must be set,

•

the

CALIBRATION TYPE

must be selected (see “CALIBRATION TYPE Menu Item”

on page 113), and

•

the

CALIBRATION DATA

must be entered (see “CALIBRATION DATA Sub-Menu”

on page 115).

NOTICE

For maximum accuracy, allow at least twenty minutes of warm-up with the exci-
tation voltage applied to the transducer before calibration.