20 Troubleshooting

20.1 Section overview

FastFacts 20.1: Troubleshooting

20.2 Self test failures

FastFacts 20.2: Self test failures

Any failure during power-on self test (P.O.S.T.) is indicated on-screen and further
use of the instrument is denied. Try the following steps to resolve the failure:

1) To access Self test, press <Setup>, then highlight and <PICK> [Self test].

2) If you received a Max out/in or a Routing failure, remove any hearing aids from

the test chamber, and rerun self test.

3) Unplug all microphones (REM & HIT). Rerun self test. If self test passes,

reconnect the microphones, one by one, running self test each time until a
failure occurs. If the failure returns, display and print the test results and
contact Audioscan for help.

This section covers common problems encountered when using the RM500SL. If
you require further assistance, note your model and serial number (located on the
recessed panel) and contact your local distributor, or Audioscan at 519-268-3313
800-265-2093 (USA only), 519-268-3256 (fax) or [email protected]

PLEASE DO NOT SHIP YOUR UNIT BACK TO THE FACTORY. In most cases
we can courier you a part that you can quickly and easily replace yourself.

This section covers the following topics:

20.2: Self test failures

20.8: Invalid HIT calibration

20.3: HIT high distortion or noise

20.9: No REM ref. mic. detected

20.4: HIT curves inconsistent

20.10: Invalid REM calibration

20.5: HIT curves differ from specifications

20.11: REM speaker overdriven

20.6: HIT speaker overdriven

20.12: Barcode scanner malfunction

20.7: No HIT reference mic. detected