Hansol Multitech Uromat 3000 User Manual

Page 59

Advertising
background image

06/00 - 59 of 90 - 0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

3.23 Operation in the Service Mode

a. Connection of a Laptop


To establish a connection with the URO controller, a PC/laptop is required.

Equipment:

⇒

PC or laptop with color display (black/white is also sufficient, but makes operation more
difficult)

⇒

486 processor, 16 MB RAM

⇒

Windows 3.x Win 95 operating system (Windows NT can cause difficulties with the Com
interface)

⇒

Hard drive

⇒

Serial connection cable (9-pole, Sub-D connector to a 9-pole, Sub-D socket, connected
1 - 1), length > 1m.

⇒

Current version of the "hp-Service" software

b. Procedure

•

Switch off the URO

•

Connect the PC and the URO CPU board using the connection cable

•

Switch on the URO

•

After approx. 1 s, the service LED will begin to blink

•

Switch on the PC

•

Start the PC software

•

Set the PC software to monitor. The terminal screen will appear

•

Set the service switch on the CPU board to the up "Service" position

•

After approx. 1 s, the service LED will switch to continuously on

•

The start message of the service menu will appear on the screen.




U U RRRR OOO 333 000 000 000
U U R R O O 3 0 00 0 00 0 00
U U RRRR O O 33 0 0 0 0 0 0 0 0 0
U U R R O O 3 00 0 00 0 00 0
UUU R R OOO 333 000 000 000


Service mode

(1) Input test
(2) Analog test
(3) Handcontrol and footswitch test
(4) Axis calibration
(5) Versioncheck

Select testmode




Advertising