:SOURce

TRIGger:SOURce BUS | EXTernal | HOLD | IMMediate

specifies the

trigger source to advance the channel list during scanning.

Parameters

Parameter Name

Parameter Type

Range of Values

BUS

discrete

*TRG or GET command

EXTernal

discrete

“Event In” BNC port

HOLD

discrete

Hold triggering

IMMediate

discrete

Immediate triggering

Comments

Enabling the Trigger Source: The

TRIGger:SOURce

command

only selects the trigger source. The

INITiate[:IMMediate]

command

enables the trigger source.

Using the

TRIG

Command: You can use

TRIGger[:IMMediate]

to

advance the scan when

TRIGger:SOURce HOLD

or

TRIGger:SOURce BUS

is selected.

Using External Trigger Inputs: With

TRIGger:SOURce EXTernal

selected, only one switchbox at a time can use the external trigger
input at the HP E1300A/E1301A mainframe “ Event In” BNC port.
The trigger input is assigned to the first switchbox that requested the
external trigger source (with a

TRIGger:SOURce EXTernal

command).

Assigning External Trigger: A switchbox assigned with

TRIGger:SOURce EXTernal

remains assigned to that source until the

switchbox trigger source is changed to

BUS

,

HOLD

, or

IMMediate

.

When the source is changed, the external trigger source is available
to the next switchbox which requests it (with a

TRIGger:SOURce EXTernal

command). If a switchbox requests an

external trigger input already assigned to another switchbox an error
is generated.

Using Bus Triggers: To trigger the switchbox with

TRIGger:SOURce BUS

selected, use the IEEE 488.2 common

command

*TRG

or the HP-IB Group Execute Trigger (

GET

)

command.

Trig Out Port Shared by Switchboxes: See the

OUTPut

command.

Related Commands:

ABORt

,

[ROUTe:]SCAN

,

TRIGger

*RST Condition:

TRIGger:SOURce IMMediate

Chapter 5

HP E1366A/E1367A RF Multiplexers Command Reference 57

Advertising
This manual is related to the following products: