HP VXI E1432A User Manual
Page 194

Constant Level Output
Output Level at 1 kHz
(after 1 second settling, amplitude scale factor >
0.001)
1 Vp (nominal)
Output Impedance
1.2 kΩ (typical)
Flatness
25 Hz to 5 kH, amplitude scale factor 0.001 to 1.0
5 Hz to 20 kHz, amplitude scale factor 0.01 to 1.0
5 Hz to 20 kHz, amplitude scale factor 0.1 to 1.0
1.13 Vp to 0.50 Vp (+10, -6.0 dB) (typical)
1.13 Vp to 0.44 Vp (+10, -7.0 dB) (typical)
1.13 Vp to 0.88 Vp (±1.0 dB) (typical)
Sine Wave Distortion
(at 1 kHz, amplitude scale factor 0.1 to 1.0)
− 40 dBc (typical)
Residual dc Offset
< 5 mV (typical)
Summer Input
Maximum Input Level
10 Vp
Gain, Summer Input to Signal Output
0 ± 0.5 dB at 1 kHz
Input Impedance
> 10 kΩ (typical)
Flatness, dc to 25.6 kHz
± 0.5 dB (typical)
Sine Wave Distortion
− 80 dBc (typical)
Residual dc Offset
1 mV (typical)