(Auto-Run, cont.)

[Summary]

Figure 8 is a confirmation screen. The Operator needs to check that the Wash
Buffer Bottle and Antibody Reservoir are filled, and that the Waste Bottle is
emptied prior to the Auto-Run. Press “Start” key. The following screens will be
displayed in order and Auto-Run will be executed:

o

Figure 9 displays the priming is in process.

o

Figure 10 displays the Rack and Pette setup is being checked.

o

Figure 11 displays the Rack and Pette setup is confirmed.

o

Figure 12 displays the washing process (Process 1, 4 or 7) or aspiration
process (Process 2 or 5) is in operation. See Chapter 6, Operation, in the AP
720S Instruction Manual.

o

Figure 13 shows incubation process (Process 3 or 6) is in operation.

o

Figure 14 is a warning screen during incubation. It indicates the door is open
with a beep sound. After the door is closed, the screen will return to Figure 13.
In the case the door is open during the operation, i.e. initialization, priming,
rack & pettes check, washing, aspirating, rinsing and draining, the screen will
switch to Pause (see section 2-8 Pause)