Honeywell 4909 CPVC User Manual

Page 31

Advertising
background image

Platinization and Platinum Black

7/99

4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual

23

WHITE

BLACK

RED

GREEN

Cell

Temp.
Comp.

W

K

SH

G

R

Conductivity/Resistivity

Analyzer

Note 1

GND

K

W

G

R

#2 Temp
Compensator

#1 Temp
Compensator

Cell 2
Electrodes

Cell 1
Electrodes

(Note 5)

(Note 5)

Five Point Terminal Board.
Each Terminal Will Accept
#16 Gage Max. Wire

An

a

l y

ze

r I np

u

t C

on

n

e

c ti o

n

s

20 ft. max.

Red
Green
Black

White

Cell cable is approx. 0.250"
(6.4mm) O.D. max. with 4

conductors of #18 AWG
wire, 7 or 20 foot length.

D

i r e

c

t C

e

l l t o

A

n

a

l y

z

e

r I n

s

t a

l l a

t i o

n

NOTES:
1. For pure water samples in non-conductive (plastic, glass, etc.) piping, ground the black cell electrode lead near the cell.
Alternatively, connect to the 9782 ground screw as shown dotted. Do not ground 10, 25, or 50 constant cells.
2. 97

8

2C

- S

0

(only)

Use 22 gage minimum coaxial cable type RG59/U connecting shield to terminal "SH" only.
3. 9

7

8

2

C

- S

0

For cable runs of up to 500 ft., use: 18 gage minimum, three conductor cable.
For cbale runs of 500 - 1000 ft., use: 16 gage minimum, three conductor cable.
9

78

2C

- W0

[coax and shield (SH) not used]

For cable runs of up to 500 ft., use: 18 gage minimum, four conductor cable.
For cable runs of 500 - 1000 ft., use: 16 gage minimum, four conductor cable.
4. Cell to analyzer cables are considered low level. Run seperate from high level wiring.
5. If 2 Cells are to be applied, the same guidelines are applied to Cell 2 as wereused for Cell 1.

An

a

l y

ze

r I np

u

t C

on

n

e

c

ti o

n

s

1000 ft. max.

Red

Green

Black

White

R

G

K
W

20 ft. max.

Coax Cable

Shield

Junction Box

Note 3

Note 2

C

e

l l t o

A

n

a

l y

z

e

r T

h

r o

u

g

h

J

u

n

c

t i o

n

B

o

x

a/n 23346

Ce

l l A

s

s

e

mb

l y

Co

n

n

e

ct

i o

n

s

4905

I

II

III

IV

V

VI

- - 333 - X1

4973
4974

I

II

III

IV

- 333 - X1 -

V

4908

I

II

III

IV

V

VI

- - 333 - X1

4909

VII

Ce

l l A

s s e

mb

l y

Co

n

n

e

ct

i o

n

s

4905

I

II

III

IV

V

VI

- - 333 - X1

4973
4974

I

II

III

IV

- 333 - X1 -

V

4908

I

II

III

IV

V

VI

- - 333 - X1

4909

VII

W

K

SH

G

R

Conductivity/Resistivity

Analyzer

Note 1

GND

K

W

G

R

#2 Temp
Compensator

#1 Temp
Compensator

Cell 2
Electrodes

Cell 1
Electrodes

(Note 5)

(Note 5)

Note 3

Figure 5-4 Installation Diagram-Cat. 4909 Cells with 7 or 20 Foot Leads

Connected to 9782 Conductivity/Resistivity Analyzer