ArbExplorer Feature Highlights

396 - Three powerful tools in one software package: complete Instrument control

panel, Waveform composer and FM signal composer

397 - Four powerful tools in one software package: complete Instrument control

panel, Waveform, pulse, and FM signal composer

397 - Detailed virtual front panels control all of the Instrument’s functions and

modes

Wave composer generates, edits and downloads complex waveforms

FM wave composer generates and downloads complex modulating signals

397 - Easy, onscreen generation of complex pulses using the pulse composer.

396 - Automatic detection of active instruments

Equation editor generates waveforms from equations

SCPI command and response editor simulates ATE operation

Translates waveform coordinates from ASCII and other formats

Simplifies generation of complex sequences

Features, Functions, and Options

The following is a general description of the features, functions, and options available

with the Instrument.

Output Channels

The 396 is a single channel arbitrary waveform generator.
The 397 is a dual-channel waveform generator. Although the two channels share a single

sample clock source, each channel can be controlled separately to generate different

waveform, amplitude and waveforms sequence. Having a single sample clock source is

an advantage in a dual-channel configuration because it allows tight control over inter-

channel synchronization and leading edge start phase where the initial skew between the

two channels is just a few nanoseconds.

Output Functions

The Instrument is completely digital. There are no analog functions resident in its

hardware circuits. Data has to be downloaded to the Instrument for it to start generating

waveforms. The Instrument can generate a few standard functions such as sine wave,

triangular wave and square wave. Each time that a standard function is required, the

Instrument calculates its coordinates and places them in the waveform memory.

Therefore, every time a standard function is selected, minimal time is required for the

controller to compute the function and load its data to the waveform memory.

Frequency

Waveform frequency is programmed with 7 digits on the 396 and 9 digits on the 397.

Frequency accuracy of the output waveform is determined by the clock reference. The

internal reference oscillator provides 1 ppm accuracy and stability over time and

temperature. If higher accuracy and/or stability are required, you may connect an external

frequency reference to the rear-panel reference input connector.