3 specifications and warranty 7, 1 specifications 7, Figure 2 standard dimensions 7 – Fluke 5626/5628 User Manual

Page 13: Table 2 specifications 7, 3 specifications and warranty, Specifications, Figure 2, Standard dimensions, Table 2, 3specifications and warranty

Advertising
background image

3

Specifications and Warranty

3.1

Specifications

See Table 2 and Figure 2.

7

3

Specifications and Warranty

5626

5628

Temperature Range

–200°C to 661°C

R

tp

100

Ω ±1Ω at 0.01°C

25.5

Ω ±0.5Ω at 0.01°C

Resistance RatioW(Ga)

W(302.9146K)

≥1.11807

α ≥ 0.003925

Calibration Uncertainty
(k=2)

±0.006°C at –200°C

±0.004°C at 0°C

±0.009°C at 420°C
±0.014°C at 661°C

Short-Term Stability

±0.003°C

±0.002°C

Long-Term Drift (R

tp

)

≤0.03°C with 500 hours at

661°C

≤0.02°C with 500 hours at

661°C

Sheath

Inconel™ 600

Handle Temperature
Range

0°C to 80°C

Lead Wires

4-wire Super-Flex PVC, 22 AWG

Termination

Gold-plated spade lugs

Size

304.8 mm x 6.35 mm dia. (12" L x 0.25")

381 mm x 6.35 mm dia.(15" L x 0.25”)

508 mm x 6.35 mm dia. (20" L x 0.25")

Table 2 Specifications

5626

SN:0021

562X-12 12"
562X-15 15"
562X-20 20”

≈2 meters

(6')

≈152 mm

(6")

Strain Relief

Terminal Box

Lead Wire

65 mm

(2.5")

Inconel Sheath

6.35 mm

(.25")

Gold-plated Terminals

Figure 2 Standard Dimensions