Problem

Causes and Solutions

Temperature readout differ-
ent than expected, e.g. the
heat source is set at 300°C,
the PRT measures 275°C.

Measure the PRT resistance at TPW.

If the resistance of the PRT is less than the rated resis-

tance, e.g. 70

Ω for the 5626, there may be a short in

the sensor. Contact an Authorized Hart Scientific Service
Center (see Section 1.3).

If the resistance of the PRT is only a few ohms, there
may be a short in the four lead-wires. Contact an Autho-
rized Hart Scientific Service Center (see Section 1.3).

If the PRT is open, the resistance will be “Out of Limits”
or in the kilo-ohm or mega-ohm range. Contact an Au-
thorized Hart Scientific Service Center (see Section 1.3).