Temperature alarm

Before installing alarm

1. Remove supplied 9-volt back-up battery from packing

box.

2. Remove 2 screws from module face and remove

faceplate.

3. Install back-up battery on battery connector.

4. Locate DIP switches on the back of the faceplate

(Fig. 13).

5. Review the factory DIP switch settings (Fig. 14) and

make any changes required to meet the needs of your
specific application.

6. Reinstall faceplate.

Dip Switch

OFF (factory default setting)

ON

1

No alarm delay

45 minute alarm delay

2

Auto reset (alarm stops automatically as

Manual reset of alarm (user must press

soon as temperature returns to set range)

RESET button to stop audible alarm even if

temperature has returned to set range)

3 & 4

Audible alarm on (no snooze with both off)

No audible alarm

3

Enables 5 min alarm “snooze”

Enables 45 min alarm “snooze”

when reset button is pressed (3 off with 4 on)

when reset button is pressed (3 on with 4 off)

Fig. 14

Fig. 15

Installing alarm

1. Mount alarm in desired location with screws (supplied

by others) through back of housing. Do not place
undue strain on probe cable.

Note: Do not modify length of probe wire. Probe will

not measure temperature correctly if wire length is
changed.

2. Plug power cord into 110 outlet.

3. Push center tab of short bottle bracket into top of rear

left pilaster (Fig. 15).

Note: Use longer bottle bracket for REF5 refrigerators

using cut-out upper shelf accessory.

4. Fill bottle with glycerin or other liquid to increase

accuracy of readings by simulating the internal
temperature of medications.

5. Insert bottle into bottle bracket.

6. Route probe through hole in refrigerator back and push

probe down through gasketed bottle top.

Note: Alarm probe must be placed in bottle for
proper system operation. Refer to Fig. 16 if attaching
alarm to central monitoring or central alarm system. A
SPDT 1 amp 24V AC resistive relay is provided for this
connection.