4 ce comments, 1 emc directive, 2 immunity testing – Fluke 1523 User Manual

Page 13: 5 using clamp-on ferrites, 1 emc directive 1.4.2 immunity testing

Advertising
background image

Before You Start

Using Clamp-On Ferrites

5

CE Comments

1.4

EMC Directive

1.4.1

Hart Scientific’s equipment has been tested to meet the European Electromagnetic Compatibility Directive

(EMC Directive, 2004/108/EC ). The Declaration of Conformity for your instrument lists the specific stan-

dards to which the unit was tested.
The instrument was designed specifically as a test and measuring device. Compliance to the EMC direc-

tive is through EN 61326-1:2006 Electrical equipment for measurement, control and laboratory use – EMC

requirements
As noted in the EN 61326-1, the instrument can have varying configurations. The instrument was tested in a

typical configuration with shielded RS-232 cables.

Immunity Testing

1.4.2

The instrument was tested to the requirements for laboratory locations.

Using Clamp-On Ferrites

1.5

Clamp-on ferrites are provided for use in improving the instrument’s electromagnetic (EM) immunity in envi-

ronments of excessive EM interference, like areas of heavy industrial equipment. We recommend placing the

ferrites on the cables of probes attached to the instrument.
To attach a ferrite to a probe cable, make a loop in the cable near the connector and clamp the ferrite around

half of the loop as shown in the diagram. The ferrite can be easily detached and moved to a new probe when

needed. (See Figure 1 on next page.)