Table 12 temperature, thermistor – Fluke 1523 User Manual

Page 36

Advertising
background image

1523, 1524 Thermometer Readout

Specifications

28

Temperature, RTD Ranges,

Table 11

and Accuracies (ITS-90)

Accuracy ±

°C 4 Wire Probe

± 0.011 at –100 °C

± 0.015 at 0 °C

± 0.019 at 100 °C
± 0.023 at 200 °C
± 0.031 at 400 °C
± 0.039 at 600 °C

Resolution:

0.001°C (0.001°F)

Temperature, Thermistor

Table 12

Accuracy ±

°C

± 0.002 at 0 °C

± 0.003 at 25 °C
± 0.006 at 50 °C
± 0.014 at 75 °C

± 0.030 at 100 °C

Resolution:

0.001 °C (0.001 °F)

Based on a 10 k

Ω (at 25 °C) thermister

with a beta value of 4000

Ω. See

technical manual for details.

Advertising