Fluke 1595A User Manual

Page 41

Advertising
background image

31

Menus and Screens

Main Menu

5.7.1.2.1 SET UP Rs (F1)

MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS |

Pressing the Set Up Rs function key displays the Rs definition assigned to each Rs input. The Up/Down arrow

keys are used to scroll through the list to select an Rs input for configuration. The following function keys al-

low the user to configure the Rs input.

5.7.1.2.1.1 ASSIGN RESISTOR (F1)

MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS | SET UP Rs |

The Assign Resistor function key opens the resistor definition library. This is a list of Rs configurations that

have been created by the user. The resistor definition library may contain up to 50 resistor definitions. Every

definition in the library must have a unique ID.
Also displayed in the resistor definition library, at the end of the list, are factory default options VARIABLE

RESISTANCE and NO RESISTOR.
The VARIABLE RESISTANCE option is used when the Rs resistance will vary such as when a SPRT is being

used as a reference resistor. When in this mode, measurement output will be in ratio only.
If the NO RESISTOR option is chosen, the Rs Input cannot be used as a reference resistor connection.

5.7.1.2.1.1.1 DEFINE RESISTOR (F2)

MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS | SET UP Rs |

The Define Resistor function key displays the resistor definition library. The displayed function keys are used

to create new Rs resistor definitions or edit definitions stored in the resistor definition library. The following

describe each function key.

5.7.1.2.1.1.2 ADD RESISTOR (F1)

MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS | SET UP Rs | DEFINE RESISTOR |

The Add Resistor screen is used to create a new Rs definition using the following configuration fields and

function key.
RESISTOR ID - used to identify the reference resistor. Pressing the ENTER key while the RESISTOR ID

field is highlighted will open the Alpha-Numeric Interface to allow entry of a unique ID.
RESISTANCE VALUE - the resistance value of the Rs reference resistor. This resistance value is used to

calculate UUT resistance so it is important that it is entered correctly.
MAXIMUM CURRENT - establishes the maximum current allowed to be applied during measurement. This

prevents excessive current from being applied to the reference resistor.
CALIBRATION DATE - used to define when the reference resistor was calibrated. If not used, the digits can

all be set to zeros.
DUE DATE - used to alert the user when the reference resistor is due for calibration. This is only applicable if

Rs DUE ALERT is enabled in the User Settings screen, see Section 5.7.4.1, USER SETTINGS (F1), on page

47.
SAVE (F1)
The Save function key saves the settings.
CANCEL (F5)
The Cancel function key exits Add Resistor without saving changes.