Menus and Screens

Main Menu

write the file. If a USB memory device has not yet been inserted, the user is instructed to do so. Allow about

10 seconds for the system to recognize the USB device before continuning.
Resistor definition files are written to a folder named “\159x\resistor” on the USB memory device and have

the extension “.res”. If the file does not exist, it will be created. If the file already exists and one resistor was

selected, its resistor definition will be appended to the data in the file. If ALL RESISTORS was selected, the

contents of the existing file will be replaced with the resistor definitions in the resistor definition library.

5.7.1.2.1.3.3

MOVE UP (F3)

MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS | SET UP Rs | MANAGE RESISTORS

The Move Up function key changes the arrangement of the reference resistors in the resistor definition library

by moving the selected resistor up one position. If the first resistor is selected or if ALL RESISTORS is se-

lected, there is no action.

5.7.1.2.1.3.4

MOVE DOWN (F4)

MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS | SET UP Rs | MANAGE RESISTORS

The Move Down function changes the arrangement of the reference resistors in the resistor definition library

by moving the selected resistor down one position. If the last resistor is selected or if ALL RESISTORS is

selected, there is no action.

5.7.1.2.1.3.5

DELETE RESISTOR (F5)

MAIN MENU | CHANNEL SETUP | CHANNEL SETTINGS | SET UP Rs | MANAGE RESISTORS

The Delete Resistor function key may be used to delete a resistor definition from the resistor definition library.
If a resistor was selected in the Manage Resistors screen, that resistor will be deleted. If ALL RESISTORS

was selected, the entire library will be erased.
A message appears prompting the user to confirm the delete operation. The user may press the YES func-

tion key to confirm and proceed with deleting the resistor definition or library. Pressing the NO function key

cancels the operation and returns to the Manage Resistors menu. Pressing the EXIT key will also cancel the

operation with no change.